Instruction manual

42
7. Maintenance 
7-1 Cleaning 
Caution 
• Before cleaning, turn off the power and remove the AC adapter. 
• Do not pour water on the instrument or use water for cleaning. This instrument is not 
waterproof. 
• Do not sterilize the instrument by gaseous sterilization or using an autoclave. 
• Do not use solvents such as thinner and benzine for cleaning. That may deform or 
discolor the instrument. 
Clean the instrument as described below: 
  Clean the instrument exterior with a dry soft cloth. 
 If the instrument is stained with blood, chemical substance, or dirt, wipe the instrument off using a cloth 
dampened with a mild detergent. 
For cleaning a patient's body site to apply a sensor, follow the procedure of each hospital.   
 To disinfect the cuff, wipe the surface which comes into contact with the skin with alcohol such as 70% 
isopropyl. Do not wash the cuff. 
  Discard the disposables after use. To disinfect the optional SpO
2
 sensor or the extension cable, use 
alcohol such as 70% isopropyl or bleach diluted by 10%. 
  Wipe the ECG extension cable and ECG patient cable with alcohol such as 70% isopropyl. 