Instruction manual

41
No.  Setting item  Setting range  Description 
S00  Station address  0-98 (0)  To connect an option. 
S10  Communication mode  1-10 (1) 
To connect an option.   
To be selected, depending on the option. 
S11 Baud rate 
1200, 2400, 4800, 
9600
, 19200, 38400
To connect an option.   
To be selected, depending on the option 
S12 Data bit  7, 8 
To connect an option.   
To be selected, depending on the option 
S13 Stream mode  OFF, ON 
To connect an option.   
To be selected, depending on the option 
S21  Previous data hold  OFF, ON 
Whether to retain the previous data or not 
when the POWER switch is turned off. 
"ON" to retain. 
S22  Averaging pulse  1, 4, 8 
Sets the number of ECG pulse rate data 
for moving average.   
S23 
Just in time 
measurement 
OFF, ON
Whether to use an exact-time measurement 
for blood pressure or not. 
"ON" to use. 
S24  Sensor removal alarm  OFF, ON 
Whether to sound an alarm when the 
sensor removal is detected. 
"OFF" not to sound. 
S25  Heart rate -pulse 
AUTO
, ECG, SpO
2
, 
NIBP 
Specifies whether to display, heart rate or 
pulse rate and pulse sound. 
"AUTO" to sound pulse in synchronization 
with the pulse detected according to the 
order of precedence. 
Underlined items: Default settings 