Instruction manual

32
Menu  Sub menu  Setting range  Unit  Description 
Cycle 
OFF,
 CON, 2.5, 3, 5, 10, 
15, 20, 30, 60, 90, 120 
min 
Sets the cycles for interval 
measurement. 
NIBP 
Pressurization AUTO,
 100-260 (by 20)  mmHg
Sets the pressurization value for 
blood pressure measurement. 
"AUTO": Detects the pulse at 
the start of pressurization to 
pressurize properly. 
During interval measurements, 
"AUTO" is used. 
Upper OFF, 50-250 (by 10) 
SYS 
Lower OFF,
 30-200 (by 10) 
mmHg
Upper OFF, 50-220 (by 10) 
Limits 
HR/PR 
Lower OFF,
 30-150 (by 10) 
bpm 
Sets the limit values for alarm.
Gain x1/2, x1, x2, x4 
⎯ 
Sets the display gain of the 
ECG waveform. 
ECG 
Time base  6.25.12.5, 25.0
 mm/s
Sets the sweep rate of the 
ECG waveform. Also applied 
to the plethysmograph. 
Underlined items: Default settings 
6-5  Measuring the Blood Pressure 
Caution 
If the instrument malfunctions during measurement, for example, the air is not 
exhausted, remove the cuff from the patient immediately. 
Blood pressure measuring procedure 
1.  Press the NIBP START/STOP key to start a measurement.   
2. When "AUTO" is selected for pressurization, the cuff pressure is adjusted automatically to the 
patient's blood pressure value. Otherwise, the cuff is pressurized to the set value. 
3. After measurement, the air is rapidly exhausted from the cuff and the measurement values are 
displayed. When the pulse rate by the ECG or SpO
2
 unit is not available, the pulse rate obtained by 
blood pressure measurement is displayed.   
4.  When the blood pressure can not be detected, the measurement is repeated three times. 
Note 
• Make sure that the cuff is at the same level as the patient's heart. 
• Advise the patient to sit comfortably and relax, remain still and keep quiet during measurement.   
• To cancel a measurement that is in progress, press the NISB START/STOP key.   
•  When an error occurs during measurement, the error code appears in the SYS display and the error 
message appears at the bottom of the display. 