Instruction manual

30
Continued 
Gain x1/2, x1
, x2, x4 
⎯ 
Sets the display gain of the 
ECG waveform. 
Time base  6.25.12.5, 25.0 mm/s
Sets the sweep rate of the 
ECG waveform. Also applied 
to the plethysmograph. 
Lead l, ll, lll  Sets the ECG lead. 
Hum filter  OFF, ON 
⎯ 
Set the ON/OFF of the hum 
filter of the ECG unit. 
Return time  0.5, 3.2  s 
Sets the time constant of the 
ECG unit. 
ECG 
Pacemaker ON, OFF
⎯ 
"ON": Highlights the pacemaker 
in the ECG waveform and 
prevents mis-counting of the 
heart rate by the pacemaker. 
"OFF" is selected upon power-on.
Upper ECG, SpO
2
Sets the waveform displayed 
at the upper area. 
Waveform 
Lower ECG, SpO
2
⎯ 
Sets the waveform displayed 
at the lower area. 
Year 2000-2099 
Month 1-12 
Day 1-31 
Hour 0-23 
Clock 
Minute 00-59 
  Sets the clock. 
Alarm sound  1-4 (2)  Sets the alarm buzzer volume
Click sound  0-4 (2)  Sets the key click volume 
Volume 
Pulse sound  0-4 (2
) 
⎯ 
Sets the pulse detection 
buzzer volume. 
Storage cycle 
OFF,
 1, 2, 2.5, 3, 5, 10, 
15 
min 
Sets the memory interval to 
store the values except blood 
pressure. 
Blood pressure values are 
stored each time they are 
measured, regardless of this 
setting. 
Temp. probe  OFF, 1-10 
Sets the temperature probe 
ID used. 
Display ON, OFF 
⎯ 
"OFF": Does not display the 
measurement values. 
Brightness 10-100% (60%) (by 5)  % 
Sets the brightness of the 
backlight. 
System 
Auto print  OFF, list, graph, trend 
⎯ 
Sets the auto printing. 
Underlined items: Default settings 