Instruction manual

29
5. Press the 
 or    key to change the value. 
6.  The changed value is confirmed when: 
  Moved to the next menu or sub menu. 
  The HOME key is pressed to return to the home 
display. 
Note 
• 
In the setting mode display, after about 5 minutes of inoperation, the home display will be indicated. 
• 
When setting the clock, note the following. 
The time in the clock menu indicates that of when the clock menu is displayed, and it remains the same 
as long as the clock menu is displayed. The actual clock of the instrument continues normal operation.   
Changing any item of the clock menu is reflected and resets the clock. 
• When an alarm occurs, the display will return to the home display. 
Setting items 
Menu  Sub menu  Setting range  Unit  Description 
Upper  OFF, 50-250 (by 10),  200
SYS 
Lower  OFF, 30-200 (by 10),   70
Upper  OFF, 30-200 (by 10),  150
DIA 
Lower  OFF, 10-200 (by 10),   40
mmHg
Upper  OFF, 50-220 (by 10),  170
HR/PR 
Lower  OFF, 30-150 (by 10),   40
bpm 
Upper OFF, 75-99 (by 1) 
SpO
2
Lower  OFF, 50-99 (by 1), 85
% 
Upper OFF, 30-44 (by 0.5) 
Limits 
Temp 
Lower OFF,
 26-40 (by 0.5) 
°C 
Sets the limit values for alarm.
Note 
When the lower limit value for 
SpO
2 
measurement has been 
set to be less than 85%, it will 
be reset to the default setting 
upon power-on. 
Cycle 
OFF,
 CON, 2.5, 3, 5, 10, 
15, 20, 30, 60, 90, 120 
min 
Sets the cycles for interval 
measurement. 
Pressurization AUTO, 100-260 (by 20)  mmHg
Sets the pressurization value for 
blood pressure measurement. 
"AUTO": Detects the pulse at 
the start of pressurization to 
pressurize properly. 
During interval measurements, 
"AUTO" is used. 
Patient Adult, Neonate 
⎯ 
"Adult": Other than neonates 
NIBP 
Notice buzzer  OFF,
 ON 
⎯ 
"ON": Sounds a buzzer 15 
seconds before starting an 
interval measurement. 
Underlined items: Default settings 