Instruction manual

25
6. Operation 
6-1  Turning on the POWER switch 
Turn the POWER switch on. After several seconds of self-check, the home display appears. 
Note 
• When the self-check detects any error, the error code is displayed. The error code display of some 
errors is frozen. Refer to "7-6 Error Codes" for how to solve the error. 
•  The self-check checks the data in memory. The duration of the self-check depends on the number of 
memory data. 
6-2  Attaching the Cuff/ECG Electrodes/SpO
2
 Sensor 
Consider "Safety Precautions" appearing in an earlier part of the instruction manual, when attaching the 
Cuff/ECG electrodes/SpO
2
 sensor. 
Attaching the cuff 
1. Hold the cuff with the hose pointing 
toward the lower part of the arm.   
2. Wrap the cuff around the upper arm, 
about 3 5 cm above the elbow, with the 
ARTERY mark on the artery, as shown in 
the illustration. 
3. Allow some slack between the cuff and 
arm so that a finger or two can be 
inserted. 
4. When the cuff inner end is not within the 
area labeled RANGE, the cuff size is not 
appropriate. Replace the cuff. 
Note 
• Constriction of the upper arm, caused by rolling up a shirt sleeve, may prevent accurate readings. 
•  When the size of the supplied cuff does not fit the patient, refer to "8-1 Accessories/Options" to find the 
proper sized cuff. 