Instruction manual

12
Charging the battery 
 Charging starts when the AC adapter is connected to the instrument, regardless of the power switch 
status. It takes 15 hours to complete charging. 
 The battery capacity lowers due to self-discharge. To compensate the capacity lost due to self 
discharge, intermittent charging is performed automatically. 
 When charging a battery with small charging capacity, supplementary charging of about one hour is 
performed. 
 Charging status is indicated by the charging lamp. 
Charging lamp  Charging status  Description 
Charging of 15 hours is being 
performed. 
Right after the AC adapter is connected.
Normal condition 
Lit continuously. 
Intermittent charging of 4 hours is being 
performed. 
Occurs about once a week when the AC 
adapter is kept connected after 15-hour 
charging. 
Normal condition 
Flashes every ten 
seconds 
Supplementary charging is being 
performed. 
Occurs in the early stage of 15-hour 
charging. 
Normal condition when the lamp 
stops flashing and is lit after a 
maximum of 4 hours. 
Flashes every two 
seconds 
Error 
Remove the battery once and 
install again. If the condition 
persists, replace the battery. 
Not lit  Charging is complete.  Normal condition. 
Note 
•  When the operating hours of the battery are short even after charging, replace the battery. Refer to "8-1 
Accessories/Options" for ordering information. 
• Replace the battery every two years. 
• The battery may have deteriorated before two years of use, due to the environment and usage 
conditions. 
• For battery operation, refer to "6-12 Power Supply". 
•  When the charging lamp will not light right after the AC adapter is connected, remove the battery and 
install it again. 