Instruction manual

5
* Available only when the optional thermometer and temperature receiver unit are connected. 
2-3 Measurement 
Blood pressure 
Measurement method: Oscillometric method 
Pressure detection method: Semiconductor pressure sensor 
Pressure range: 0 - 300 mmHg 
Pressure accuracy: ± 3 mmHg       
Measurement range: Blood pressure: 10 - 280 mmHg 
Pulse rate: 30 - 200 bpm (Neonate: 30 - 240 bpm) 
Pressurizing method: Diaphragm pump 
Pressure control method: ECEV 
Rapid air exhaust system: Electromagnetic valve 
Safety mechanism: The  electromagnetic valve opens when approx. 320 
mmHg or greater is detected. 
  For neonates, the electromagnetic valve opens when 
160 mmHg is detected. 
Degree of protection against electrical shock: Type BF 
Blood pressure 
interval measurement 
Intervals: CON, 2.5, 3, 5, 10, 15, 20, 30, 60, 90, 120 min.   
Maximum length of CON measurement is 5 minutes. After five minutes, 
the interval is switched to 5-min interval automatically. 
SpO
2
Measurement method: 2-wavelength pulse oximetry 
Measurement range: 40 - 100% 
Accuracy: ± 3 %(70-100%)   
Pulse rate measurement range: 30 - 250 bpm      
Degree of protection against electrical shock: Type BF 
Patient input 
Lead: 3-lead (Lead l, ll, lll) 
Input impedance: 5 MΩ or greater 
Input circuit current: 10
-7
 A or less 
Differential dynamic range: ±300 mV 
Degree of protection against electrical shock: Type CF 
ECG 
Measurement 
Frequency characteristics: 0.3 - 60 Hz (when 0.5 second of time constant 
is selected) 
  0.05 - 200 Hz (when 3.2 seconds of time 
constant is selected) 
Hum elimination: 50/60 Hz 
Measurement range: 20 - 300 bpm 
Caution 
For safety and best performance, use the specified accessories and options. 
If ECG electrodes other than specified are used, use much care. A great offset may 
occur. 