Error Codes
Caution The error code may be updated without prior notice.
Error
code
Meaning Status Operation and Treatment
E00
No clock parameter All parameters are lost.
Reset status.
Enter clock parameters. Refer
to "Parameters for the Display
and Clock"
E03
Pressure zero error An error code is displayed
without cuff inflation.
Release the air from the cuff
completely.
E04
Low battery Measurement is stopped.
An error code is displayed.
Auto mode is quit.
Replace with new batteries.
Restart the auto mode if you use it.
E05
Inflation error Inflation pressure does not
reach the target pressure.
Connect the cuff to the main unit
securely. If you cannot clear the
error, there may be an air leak and
repair is necessary
E06
Above 320 mmHg An error code is displayed. Do not move and try to relax
during the measurement. If you
cannot clear the error, repair is
necessary
E07
Controlled stop
using STOP key
Air is exhausted. An error
code is displayed.
Press the STOP key only when
necessary.
E08
Pulsation cannot be
measured
Measurable pulsation is
searched to 20mmHg in
constant exhaust. An error
code is displayed.
Do not move and try to relax
during the measurement. The
error occurs when pulsations
are not detected due to thick
cloth or quick motion.
E10
Pulsations cannot
be detected
because the patient
may have moved.
In the measurement, quick
exhaust is executed. An
error code is displayed.
Do not move and try to relax
during the measurement.
E20
Pulse rate < 30
200 < Pulse rate
An error code is displayed. Measure the blood pressure by
other methods.
E21
DIA < 40
160 < DIA
E22
SYS < 60
280 < SYS
E23
DSD < 10
150 < DSD
DIA : Diastolic Blood Pressure
SYS
: Systolic Blood Pressure
DSD
: The Difference between Systolic Blood Pressure
:and Diastolic Blood Pressure.