Instruction manual
Page 36
Cleaning the Cuff and Recorder
Before cleaning the recorder, remove the battery cover, turn the power switch off
and remove the batteries.
The recorder is not water resistant. Do not allow liquids to splash on or get into the
case while cleaning.
After each use, wipe the case of the recorder with a clean lint free cloth, moistened
with water and a mild detergent.
Do not use antiseptic solutions, Alcohol, etc., to clean the recorder, hose or cuff.
Clean the cuff cloth and cuff cover by washing in water with a mild detergent. Do
not scrub or wring them by hand. If the cuff cloth and cuff cover become
contaminated, replace them with new ones.
Periodical Inspection
This blood pressure recorder is a precision instrument. Please inspect the
functions (once a year) periodically. Contact your nearest A&D dealer for the
inspection.
Confirm the accuracy of the blood pressure monitor once a year.
Troubleshooting
Caution
Do not open the case of the recorder because it uses delicate electrical
components and intricate air unit that could be damaged.
If you cannot locate and fix the problem, request service from your nearest A&D
dealer, or from the A&D service group.
A&D service group will support authorized suppliers about technical information,
spare parts and units.
Problem Cause Treatment
No display at turning
on.
Batteries are drained. Replace with new batteries.
Data lost while
replacing batteries.
The built-in backup
battery is drained
Turn on the recorder and
display the clock for 24 hours
or more, to fully charge the
backup battery.
No pressure. Air leakage at the
connector, hose or cuff.
Confirm the cuff and air hose
are not damaged and are
connected correctly.