Instruction manual
Page 29
Operation
Automatic Measurement
Caution
Automatic measurement uses the internal clock and parameters for automatic
measurement. Refer to "Parameters for the Display and Clock" and "Selection
for the Automatic Measurement" for setting these parameters.
When the patient stops the automatic measurement or removes the cuff, press and hold
the AUTO ON/OFF key for approx. 3 seconds to turn off the "A" sign. If the "A" sign is left
turned on, the recorder may start the next automatic measurement and cause damage to
the cuff.
Starting or re-starting automatic measurement
Step 1 Confirm the parameters for automatic measurement. Refer to "Selection for the
Automatic Measurement".
Step 2 Press and hold the AUTO ON/OFF key for about 3 seconds. Then "A" is displayed and
the recorder starts an automatic measurement based on the internal clock and the
parameters for automatic measurement.
Operation using mode II
Step 1 Press the AUTO ON/OFF key to turn off "S" when the patient wakes up.
Step 2 Press the AUTO ON/OFF key to turn on "S" when the patient goes to bed.
Stopping or canceling automatic measurement
Step 1 Press and hold the AUTO ON/OFF key for about 3 seconds. "A" turns off and the
recorder stops automatic measurement.
Manual Measurement
Step 1 Press the START STOP key. The recorder starts a measurement. The results are
displayed and stored in memory.
To Stop the Current Measurement
Step 1 Press the START STOP key during measurement. The recorder will stop the
measurement and releases the air from the cuff.