User manual
A&D Board User Manual and Test Procedure
15. [ ] Turn the TDS210 SEC/DIV knob clockwise, to display approximately 4 waveforms. See Figure 9.
Figure 9: Oscilloscope Display of 40 kHz Digital Clock
16. [ ] Verify that the Digital Clock is producing a square wave, with about 5V Pk-Pk and 40 kHz Frequency.
The waveform should rise and fall quickly. The tops and bottoms may have some ringing, undershoot, and
noise, which is normal.
Sine/Square Wave Function Generator
1. [ ] Remove the red and black mini-grabbers from the 10kΩ resistor.
2. [ ] Connect a 10k
Ω resistor between A&D Board GEN and GND socket connectors.
3. [ ] Connect the scope probe mini-grabbers across the 10k
Ω resistor, black to GND.
4. [ ] Turn the A&D FREQ and FINE FRQ knobs fully counter-clockwise.
5. [ ] Set the A&D OFFSET and AMPLITUDE knobs to mid-position.
6. [ ] Set the A&D SHAPE toggle to “SIN” for Sine Wave output.
7. [ ] Set the A&D OFFSET SW to “OFF”.
8. [ ] Press AUTOSET.
9. [ ] Turn the TDS210 CH1 VOLTS/DIV knob counter-clockwise to indicate 2.00 V/division.
v. 2.2, © 2005-2006 by R. B. Lineberry, W. C. Headley, & R. W. Hendricks p. 15