15. [ ] The oscilloscope should display a straight horizontal line at about 5V. See Figure 5.
Figure 5: Oscilloscope Display of +5V Power Supply
16. [ ] Verify that the Mean voltage is 5V ± 5%.
17. [ ] Verify that the Pk-Pk voltage is around 300 mV or less.
18. [ ] Remove the red and black mini-grabbers from the 10k
Ω resistor.
19. [ ] Connect the 10k
Ω resistor between +9V and GND.
20. [ ] Turn the TDS210 CH1 VOLTS/DIV knob counter-clockwise to indicate 5.00 V/division.
21. [ ] Connect the scope probe mini-grabbers across the 10k
Ω resistor, black to GND.
22. [ ] The oscilloscope waveform should be a straight horizontal line at about +9V. See Figure 6.
Figure 6: Oscilloscope Display of +9V Power Supply
23. [ ] Verify that the Mean voltage is +9V ± 5%.
24. [ ] Verity that the Pk-Pk voltage is around 400 mV or less.
25. [ ] Remove the red and black mini-grabbers from the 10k
Ω resistor.
26. [ ] Connect the 10k
Ω resistor between -9V (minus 9V) and GND.
v. 2.2, © 2005-2006 by R. B. Lineberry, W. C. Headley, & R. W. Hendricks p. 12