Example of calculation:
In the case of designing a scale with a capacity of 60 kg and minimum division of 0.005
kg, using one load cell with a rated capacity of 100 kg and rated output of 1 mV/V:
Rated capacity of Load cell: A = 100 kg
Rated output: B = 1 mV/V
Minimum division: d = 0.005 kg
Excitation voltage: E = 5000 mV
Number of load cells: n = 1
Therefore, there is no problem with this design.
5-4. Setting Up the Indicator
After confirming that nothing is touching the key switches on the indicator, connect
the main power cable to a power source that has an earth ground terminal.
5-5. Parameter Settings and Calibration
When a platform scale is initially set up, it is necessary to set some parameters and
calibrate it using a calibration weight.
See “8. SETTING FUNCTIONS OR UNITS FOR USE” to set parameters and “7.
CALIBRATION” to calibrate the scale.