Specifications

19
10.4.
Restoring The Calibration Data To The Factory Set Values
If the gravity acceleration value or calibration data is changed unintentionally, restore
those values to the factory set values, as follows.
1. Refer to "
10.1. Calibration Mode" to enter the
calibration mode. The gravity acceleration value
is displayed.
2.
Tap the UNITS switch twice to display
.
3.
Tap the TARE switch to display
with " "
blinking.
4.
Tap the SAMPLE switch.
changes to
with " " blinking.
To cancel the restoring procedure, tap the
switch. The display returns to step 2.
5. When
is displayed, tap the TARE switch.
The factory set values are restored and
is
displayed. Then, the display returns to
.
To finish the setting procedure,
touch the
switch or press the
CAL switch. The scale turns the
power OFF.
Tap
Tap
Tap
Tap
twice