Specifications

14
8.2. Entering The Comparator Values
1. In the weighing mode, touch and hold the UNITS switch to enter the comparator
value setting mode.
2. Enter the comparator values using the following switches.
UNITS To shift the blinking digit to the right.
SAMPLE To increase the value of the blinking digit by one.
TARE To store the setting value and to go to the next limit setting or .
The minus sign can be set at the next digit of the least significant digit.
The SAMPLE switch alternates the minus sign on and off. The blinking "-" shows
minus and no sign shows plus.
3. When the setting is complete, the scale memorizes the values and displays ,
then returns to the weighing mode. (At this time, power-on-zero is not performed.)
Five-level Three-level Seven-level
comparator mode comparator mode comparator mode
Touch
& hold
Touch
& hold
Touch
& hold
Tap
Tap
Tap
Tap
Tap
Tap
Tap
Tap
Tap
Tap
Tap
Tap
Weighing
mode
To set the
LOLO
limit value
To set the LO
limit value
To set the HI
limit value
To set the HIHI
limit value
Weighing
mode
Weighing
mode
Weighing
mode
Weighing
mode
Weighing
mode
To set the LO
limit value
To set the HI
limit value
To set level 1
lower limit value
To set level 2
lower limit value
To set level 3
lower limit value
To set level 3
upper limit value
To set level 4
upper limit value
To set level 5
upper limit value