Product specifications
Page 84 AD-4402
8.3.
8.3.8.3.
8.3. Built
BuiltBuilt
Built-
--
-in Current Loop
in Current Loopin Current Loop
in Current Loop Output
Output Output
Output
Transmission system EIA RS-232C, Asynchronous, bi-directional, half-duplex
Current 1 = 20mA, 0 = 0 mA, external DC current source
Data length 7 bits
Start bit 1 bit
Parity bit Even
Stop bits 1 bit
Baud rate 600 bps, 1200 bps, 2400 bps
Code ASCII code
8.3.1.
8.3.1.8.3.1.
8.3.1. Connection
ConnectionConnection
Connection
The current loop output has no polarity.
Use an external DC current source.
Connect the FG terminal when using a shielded cable.
8.3.2.
8.3.2.8.3.2.
8.3.2. Communication Modes
Communication ModesCommunication Modes
Communication Modes
There are the following modes.
Stream Mode
The data is output at every display update. If the data can not be output completely
due to a slow baud rate, the data is output at the next update.
Auto Print Mode
The data is printed at batch finish and recipe finish automatically.
Manual Print Mode
When the preset print key is pressed or terminal is connected, data is output.
Total Print Mode
When accumulating it or canceling the last result, the results of batch finish and recipe
finish are printed. When canceling the last results, the inverse polarity data is output.
Current loop Output
RS-485
A
B
SG
C.LOOP
FG
Shield
Current loop Output
AD-44
02
inside
The CURRENT
LOOP terminals
do not have
polarity.