Product specifications
Page 80 AD-4402
8.2.5.
8.2.5.8.2.5.
8.2.5. A&D Data Format
A&D Data FormatA&D Data Format
A&D Data Format
This format is used for stream mode, auto print mode and total print.
This format is compatible to the AD-4325 indicator.
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18
S T , N T , + 0 0 1 2 . 3 4 k g
CR LF
Header 1
ST Stable
US Unstable
LO Out of range
Header 2
GS Gross value
NT Net value
TR Tare value
Weighing data
Data uses BCD code, is 7 figures and includes a decimal point.
When data is negative, a minus sign is appended to the head.
When data is out of range, all numerical characters are space (20h).
Unit
kg, g or t
Terminator
CR or CR + LF CR: 0Dh, LF: 0Ah
8.2.6.
8.2.6.8.2.6.
8.2.6. Address
AddressAddress
Address
Set the address in [r5 f- 8].
32 indicators can be connected to a computer.
There is no relation between communication mode [r5 f- 2] and address [r5 f- 8].
Broadcast Address
When address @oo is used, send a command to all indicators at the same time.
Example:
@ 0 0 C T A R
CR LF
Header 1
Header 2
Weighing data
Unit
Terminator
Address