Product specifications

Page 72 AD-4402
Kind No. Description
When the displayed value can not be set to zero with re-zero or tare,
the message is displayed.
0 Dispaly can not be zeroed by zero compensation.
Zero error
ZR.ERR
1 Dispaly can not be zeroed by tare operation.
When the weighing value is out of range and emergency stop is
performed, this symbol is displayed.
1 Weighing value is out of range.
Alarm 1
ALARM 1
9 Emergency stop has been performed.
It can not weigh. Check the weighing system.
Example: loadcell cable, connctors.
1
A/D converter is positve over count.
Check the loadcell cable.
2
A/D converter is negative over count.
Check the loadcell cable.
Alarm 2
ALARM 2
4
RAM error.
Check the backup battery