Product specifications
AD-4402 Page 105
10.
10.10.
10. Function List
Function ListFunction List
Function List
The function list stores parameters to control the indicator.
The parameters are stored in an item even without power supplied.
An item is classified by a category address, and is further classified by an item
number.
Refer to " 10.1.2. Outline of the Function List".
The category address has a symbol for the 7-segments display
There are two kind of the function modes to operate the function list.
Parameter settings This mode is used to change the parameter.
Referring parameters This mode is used to refer the parameter in the
weighing sequence.
Example of an item:
Category address:
[Function] - [Function setting] - [General] - [Weighing]
genf- 1 Display refresh rate
1 5 times / second
Caution
When entering Parameter settings of the function mode, the current
weighing sequence is stopped.
10.1.1.
10.1.1.10.1.1.
10.1.1. Operation Keys
Operation KeysOperation Keys
Operation Keys
To enter the function list Press and hold the ENTER key and press the
key in the weighing mode.
Select the menu Function using the key and
the ENTER key.
To select the parameter address The , SHIFT + , ENTER, ESC keys.
To change the parameter The , SHIFT + , Alphanumerical, ENTER,
ESC keys.
To store it and exit the function list The ESC key.
(To return to weighing mode)
Item number
Category address symbol
Meaning of the address
An Item
Parameter
Address
Meaning of the parameter