Product specifications

Page 100 AD-4402
9.3.4.
9.3.4.9.3.4.
9.3.4. Testing the A/D Converter
Testing the A/D ConverterTesting the A/D Converter
Testing the A/D Converter
The A/D converter data is displayed.
When pressing the ENTER key, a test voltage can be input to the A/D converter.
9.3.5.
9.3.5.9.3.5.
9.3.5. Testing the BCD Output of OP
Testing the BCD Output of OPTesting the BCD Output of OP
Testing the BCD Output of OP-
--
-01
0101
01
Tests the output of the terminals. An active output of level "1" shifts for each terminal.
9.3.6.
9.3.6.9.3.6.
9.3.6. Testing the Relay Output of OP
Testing the Relay Output of OPTesting the Relay Output of OP
Testing the Relay Output of OP-
--
-02
0202
02
Tests the output of the terminals. An active output of level "1" shifts for each terminal.
9.3.7.
9.3.7.9.3.7.
9.3.7. Testing the RS
Testing the RSTesting the RS
Testing the RS-
--
-422/485 Interface of OP
422/485 Interface of OP422/485 Interface of OP
422/485 Interface of OP-
--
-03
0303
03
When pressing the ENTER key each time, the test data "ST,GS,+0000000kg CR LF"
is output.
16.001kg
A/D Count L/C Output(mV/V)
964583 3.103872
Int.Count Test Input
160008 0ff
A/D converter output
Internal count
Loadcell output
Test input
12345kg
OP-01 000000000000000000
a18 a1
000001000000000000
b18 b1
A1
B1
A18
B18
123kg
OP-02 0001 00000
10 7 5 1
Output terminal
1 ~ 5
Output terminal
7 ~ 10
Sent data
123 kg
RNET
ST,GS,+0000000kg
←↓
0p-03 rcu and 5nd
Received data
Parity: p
Framing:
f