Product specifications
7
Class Item and Parameter Description
0
No pause
pU5e
Data output pause
1
Pause (1.6 seconds)
Selects the data output
interval.
0
Not used
at-f
Auto feed
1
Used
Selects whether or not
automatic feed is performed.
0
No output
1
AD-8121 format
info
GLP output
2
General data format
Selects GLP output method.
For how to set time and date
to be added, refer to the
balance instruction manual,
"Clock and Calendar
Function".
0
Not used
dout
Data output
ar-d
Zero after output
1
Used
Adjusts zero automatically
after data is output
0
600 bps
1
1200 bps
2
2400 bps
3
4800 bps
4
9600 bps
bp5
Baud rate
5
19200 bps
0
7 bits, even
1
7 bits, odd
btpr
Data bit, parity bit
2
8 bits, none
0
CR LF
Crlf
Terminator
1
CR
CR: ASCII code 0Dh
LF: ASCII code 0Ah
0
A&D standard format
1
DP format
2
KF format
3
MT format
4
NU format
type
Data format
5
CSV format
Refer to the balance
instruction manual,
"Description of Item "Data
Format".
0
No limit
t-Up
Timeout
1
1 second
Selects the wait time to
receive a command.
0
No output
erCd
AK, Error code
1
Output
AK: ASCII code 06h
0
Not used
5if
Serial
interface
Ct5
CTS, RTS control
1
Used
Controls CTS and RTS.
0
Water temperature
d5 fnc
Density function
ldin
Liquid density input
1
Liquid density
Available only when density
mode is selected
mlt
Programmable-unit (Multi-unit)
Available only when programmable-unit mode is selected.
Refer to the balance instruction manual, "Programmable Units"
for details".
Unit
Unit
Refer to the balance instruction manual, "Weighing Units".
: Functions for GX-K series.
:
Factory settings. Digit is a unit of minimum weighing value.
Caution
The balance may not transmit the data completely at the specified refresh rate, depending
on the baud rate or data added to the weighing data such as time, date and ID number.