Product specifications
5
3-2. Details Of The Function Table
Class Item and Parameter Description
0
Fast response, sensitive value
FAST
1
MID.
Cond
Condition
2
Slow response, stable value
SLOW
Can be changed by response
adjustment. With "
Hold 1
", sets
the averaging time.
0
Stable when within
±
1 digit
1
5t-b
Stability band width
2
Stable when within
±
3 digits
The stabilization indicator
illuminates with the display
fluctuation within the range. With
"
Hold 1
", sets the stable range.
0
OFF
Hold
Hold function
1
ON
Holds the display when stable in
animal mode. With "
Hold 1
",
ANIMAL
turns on.
0
OFF
1
Normal
2
Strong
trc
Zero tracking
3
Very strong
Keeps zero display by tracking zero
drift.
0
5 times/second
5pd
Display refresh rate
1
10 times/second
Period to refresh the display
0
Point (.)
pnt
Decimal point
1
Comma (,)
Decimal point format
0
OFF
p-on
Auto display-ON
1
ON
Turns on the weighing mode
display when the AC adapter is
connected.
0
OFF
poff
Auto display-OFF
1
ON (10 minutes)
Turns off the display after 10
minutes of inactivity.
0
OFF
g5i
Capacity indicator
1
ON
Capacity indicator.
Zero: 0%
Maximum capacity: 100%
0
OFF
add
Accumulation function
1
ON
Displays and outputs the total
value of the weighing data.
0
Displays
ba5fnc
Environment
Display
rng
Display at start
1
Does not display
Select whether or not to display
the smallest displayable weighing
value at weighing start.
Cl adj
Clock
Refer to the balance instruction manual, " Clock and
Calendar Function".
Confirms and sets the time and
date. The time and date are
added to the output data.
0
No comparison
1
Comparison, excluding "near zero" when stable value or overloaded
2
Comparison, including "near zero" when stable value or overloaded
3
Continuous comparison, excluding "near zero"
Cp
Comparator mode
4
Continuous comparison, including "near zero"
0
Sets the upper / lower limit
value
Select
Cp Hi
or
Cp lo
.
Cp in
Data input method
1
Sets the reference value
Select
Cp ref
or
Cp lmt
.
0
Not added
Cp-r
Comparison results
1
Added
Select whether or not to add the
comparison results to the output
data.
0
OFF
Cp-b
Main display comparison
1
ON
Displays the results on the main
portion of the display in place of
the weight value.
0
OFF
bep
LO buzzer
1
ON
Select whether or not to sound
the LO buzzer.
0
OFF
bep
OK buzzer
1
ON
Select whether or not to sound
the OK buzzer.
0
OFF
Cp fnc
Comparator
bep
HI buzzer
1
ON
Select whether or not to sound
the HI buzzer.
: Functions for GX-K series.
:
Factory settings. Digit is a unit of minimum weighing value.
Displayed
only when
Comparator
output