Product specifications
3
Comparator output
Contact-outputs, output the comparison results between the weighing data and
upper/lower limit values, using HI, OK, and LO.
Whether or not to sound a buzzer, depending on the results, can be selected.
Analog output
Two modes are available: To convert the specified weight value digits to voltage, and to
convert the weight value, in the range from zero to the weighing capacity, to voltage.
Output voltage range selection: Using the slide switch (0V~/0.2V~) located on the option
panel, the output voltage range can be switched between 0 V to 1 V and 0.2 V to 1V. The
default setting at shipment is 0 V to 1 V.
RS-232C
The RS-232C interface is used to communicate with a printer or a personal computer. Using
the RS-232C interface, the following operations are available through a command from the
computer:
Outputs the weighing data. Enter balance settings.
Controls the balance. Reads the balance settings.
Current loop
Current loop is a data output interface, mainly used as a printer interface.
GLP output
GLP-compliant data output is available for RS-232C and current loop. Refer to the balance
instruction manual for details about GLP output.
2-1. Accessories
Each option is provided with the following accessories.
GX-04K: DIN connector (plug) 1 piece
...........
Instruction manual (this document) 1 copy
GX-06K: DIN connector (plug) 1 piece
...........
Screwdriver 1 piece
...........
Instruction manual (this document) 1 copy