Product specifications
34
6-4. Switching Output Voltage
The output voltage can be switched using the slide
switch on the GX-06K panel. "0V~" has been set at
factory before shipment.
"0V~" (0V to1 V):
At zero 0.000 V At full scale 1.000 V
"0.2V~" (0.2 V to1 V):
At zero 0.200 V At full scale 1.000 V
6-5. Output Voltage Fine Adjustment
The output voltage has been adjusted at the factory
before shipment.
Using the ZERO and SPAN fine-adjustment control
and a voltmeter, the output voltage can be fine
adjusted.
Fine-adjustment procedure
1. Turn the display off, using the ON:OFF key. At this
time, the output voltage will be at zero. Turn the
ZERO control so that the voltmeter indicates 0.000 V
when the slide switch is set to "0V~"; 0.200 V when
the slide switch is set to "0.2V~".
2. While pressing and holding the SAMPLE and
PRINT keys, press the ON:OFF key. The balance
displays p5 . At this time, a voltage of 1 V is
generated. Turn the SPAN control so that the
voltmeter indicates 1.000 V.
3. Repeat steps 1 and 2 until the correct output voltage
is obtained.
Slide switch
"0V~" "0.2V~"
GX-06K
2 pin, -
7 pin, +
SPAN control
ZERO control
Display for setting the output
to 0 V (0.2 V)
Display for setting the outpu
t
to 1 V