Product specifications

32
6-3. Function Table Of The Analog Output
The “Analog output (aout)” of the function table can be selected when the GX-06K option is
installed in the balance.
Analog output function table
Item and Parameter Description
0
2-digit output
Converts the consecutive 2 digits, with the digit selected in
5el
as the
least, to voltage and outputs.
1
3-digit output
Converts the consecutive 3 digits, with the digit selected in
5el
as the
least, to voltage and outputs.
2
Net full scale output
Outputs 0.000 V when the net weight is zero.
Outputs 1.000 V when the net weight is full scale.
Outputs 0.000 V when the display is set to zero using the
RE-ZERO
key.
an
Analog output mode
3
Gross full scale output
Outputs 0.000 V when the gross weight is zero.
Outputs 1.000 V when the gross weight is full scale.
Tare operation using the
RE-ZERO
key will not affect the output.
(Note: If the tare is extremely light, tare operation might change the
zero point, thus it will affect the output.)
Select the least digit to be output in the mode selected in
an
.
Only available when
0
or
1
is selected as the output mode.
0
Select the first digit as the least.
1
Select the second digit as the least.
2
Select the third digit as the least.
3
Select the fourth digit as the least.
4
Select the fifth digit as the least.
5el
Output digit selection
5
Select the sixth digit as the least.
:
Factory settings. Digit is a unit of minimum weighing value.