Product specifications
31
6-2. Analog Output (GX-06K) Specifications
The specifications of the analog output unit (GX-06K) are as follows:
Output impedance 100 Ω or less
Linearity ±0.3% or less
Output connector 7-pin DIN connector
Pin connections Output Pin 7
GND Pin 2
Output range 0 V to 1 V (With the slide switch set to "0V~")
0.2 V to 1 V (With the slide switch set to "0.2V~")
Input impedance of the device connected 10 kΩ or greater
Pin Assignments Panel View
Pin No. Description
1 No connection
2 Analog GND(0 V) (Analog output)
3 Sending loop
(Current loop)
4 No connection
5 Sending loop
(Current loop)
6 No connection
7 Analog output (Analog output)
Housing Shield
Circuit
Current loop output specifications
For details on the current loop, Refer to “5-2.Current Loop Output (GX-04K/GX-06K) Specifications”.
Transmission system 20 mA current loop (Passive)
Transmission Asynchronous, uni-directional (Only from the balance)
Data format Baud rate: 600, 1200, 2400, 4800, 9600 19200 bps
Data: 7 or 8 bits
Parity: Even, Odd (Data 7 bits)
None (Data 8 bits)
Stop bit: 1 bit or 2 bits
Code: ASCII
One character format
Notes
To use the current loop, an external power supply that provides 20 mA is required.
The maximum rated voltage of the current loop is 25 V.
When a baud rate of 4800 bps or higher is used, communication may not be performed properly.
Span
Zero
Slide switch
0V~ / 0.2V~
Start bit
Data bits
Parity bit
Stop bit
20m
A
0m
A
10
Ω
A
nalog GND(0 V)
2 pin
Housing
7 pin
Analog output