Product specifications
15
4-5. Example Of Use
The following example uses the AD-8951 comparator light, which is sold separately, to
display the comparison result in red, green, or orange.
1 Using the AX-KO507-W200 cable, which is sold separately, connect the comparator light to
the balance as shown below.
2. Set the “Comparator (Cp fnc)” of the balance function table as follows:
Cp 3
Compares the result excluding the data near zero continuously.
Cpin 0
Inputs the upper or lower limit values.
bep 1
Sounds the buzzer for LO.
bep 0
Does not sound the buzzer for OK.
bep 1
Sounds the buzzer for HI.
3. Set the upper and lower limit values as follows:
Cp Hi
1010.0 g (Upper limit)
Cp lo
990.0 g (Lower limit)
4. Functions of the comparator and buzzer are as
follows, depending upon the comparison result.
Weighing data Comparator light Buzzer
900.0 g Orange Sounds
1000.0 g Green Does not sound
1100.0 g Red Sounds
A
X-KO507-W200
GX-04
K
Comparator light
AD-8951
Red HI
Green OK
Orange LO