Instruction manual

52
Calibration report using an external weight
When the setting is “info 1”: When the setting is “info 2”:
Calibration test report using an external weight
(Calibration test does not perform calibration.)
When the setting is “info 1”: When the setting is “info 2”:
AD-8121 format General data format
Manufacturer
Model
Serial number
ID number
Date
Time
Calibration type
Calibration
weight
Signature
A & D<TERM>
MODEL FX-3000i<TERM>
S/N 01234567<TERM>
ID ABCDEFG<TERM>
DATE<TERM>
<TERM> *
TIME<TERM>
<TERM> *
CALIBRATE D(EXT .)<TERM>
CAL.WEIGHT<TERM>
+2000.00 g<TERM>
SIGNATURE <TERM>
<TERM>
<TERM>
----------------<TERM>
<TERM>
<TERM>
A &
 
D
MODEL FX-3000i
S/N 01234567
ID ABCDEFG
DATE 2008/12/31
TIME 12:23:34
CALIBRATED(EXT.)
CAL.WEIGHT
+2000.00 g
SIGNATURE
<TERM>
C
R
L
F
C
R
Space, ASCII 20h
Terminator, , or
Carriage return, ASCII 0Dh
Line feed, ASCII 0Ah
C
R
L
F
AD-8121 format
General data format
Manufacturer
Model
Serial number
ID number
Date
Time
Calibration test
type
Zero point value
Tar ge t w ei ght
value
Tar ge t w ei ght
Signature
A & D<TERM>
MODEL FX-300 0i<TERM>
S/N 012345 67<TERM>
ID ABCDE FG<TERM>
DATE<TERM>
<TERM> *
TIME<TERM>
<TERM> *
CAL.TEST(EXT.)<TERM>
ACTUAL<TERM>
0.00 g<TERM>
+1999.99 g<TERM>
TARGET<TERM>
+2000.00 g<TERM>
SIGNATURE<TERM>
<TERM>
<TERM>
----------------<TERM>
<TERM>
<TERM>
A & D
MODEL FX-3000i
S/N 01234567
ID ABCDEFG
DATE 2008/12/31
TIME 12:23:34
CAL.TEST(EXT.)
ACTUAL
0.00 g
+1999.99 g
TARGET
+2000.00 g
SIGNATURE
<TERM>
C
R
L
F
C
R
Space, ASCII 20h
Terminator, , or
Carriage return, ASCII 0Dh
Line feed, ASCII 0Ah
C
R
L
F
With FZ-i series, the time and
date are added to output data.
With FZ-i series, the time and
date are added to output data.