Specifications

7
3. Precautions
To get the optimum performance from the balance and acquire accurate weight data, note the
following:
3-1. Before Use
Install the balance in an environment where the temperature and humidity are not excessive. The best
operating temperature is about 20°C / 68°F at about 50% relative humidity.
Install the balance where it is not exposed to direct sunlight and it is not affected by heaters or air
conditioners.
Install the balance where it is free of dust.
Install the balance away from equipment which produces magnetic fields.
Install the balance in a stable place avoiding vibration and shock. Corners of rooms on the first floor
are best, as they are less prone to vibration.
The weighing table should be solid and free from vibration, drafts and as level as possible.
Level the balance by adjusting the leveling feet and confirm it using the bubble spirit level.
Make sure that the bubble is in the center of the red circle.
When the bubble is off to the left :
Turn the leveling foot on the front right in the
clockwise direction.
When the bubble is off to the right :
Turn the leveling foot on the front left in the
clockwise direction.
When the bubble is off to the backward position :
Turn both leveling feet on the front in the
clockwise direction at the same time.
When the bubble is off to the forward position:
Turn both leveling feet on the front in the
counterclockwise direction at the same time.
Ensure a stable power source when using the AC adapter.
Plug in the AC adapter and warm up the balance for at least one hour.
Calibrate the balance before use or after having moved it to another location.
Caution: Do not install the balance where flammable or corrosive gas is present.
Red circle
Bubble
OK NG
Bubble spirit level
U
P
Leveling foot
DOWN
DOWN
Leveling foot
UP