Specifications

69
20. Specifications
FZ-1200CT FZ-700CT FZ-500CT
1260 ct 760 ct 510 ct
Weighing capacity
252 g 152 g 102 g
1260.042 ct 760.042 ct
*
1
510.042 ct
Maximum display
252.008 g 152.008 g
*
1
102.008 g
0.001 ct
Minimum weighing value
(1 digit)
0.001 g
0.001 ct
Repeatability
Standard deviation
0.0005 g
±0.002 ct
Linearity
±0.001 g
Stabilization time
(typical at FAST)
Approx. 2 second
*
1
Sensitivity drift
(10 °C - 30 °C / 50 °F - 86 °F )
±2 ppm/°C
Internal mass
Available
Clock and calendar function
Available
Operating environment
5 °C to 40 °C (41 °F to 104 °F) 85 %RH or less (No
condensation)
Display refresh rate
Approx. 5
*
2
, 10 times/second
Minimum unit mass
0.001 g
Counting
mode
*
3
Number of samples
10, 25, 50 or 100 pieces
Minimum 100% reference m-ass
0.100 g
Percent
mode
*
3
Minimum 100% display
0.01%, 0.1%, 1% (Depends on the reference mass stored.)
Interface
RS-232C
External calibration weight
250 g
200 g
100 g
50 g
25 g
150 g
100 g
50 g
20 g
100 g
50 g
20 g
Weighing pan diameter
90 mm
External dimensions
198(W) x 294(D) x 315(H) mm
AC adapter
Confirm that the adapter type is correct for the local voltage
and power receptacle type
Power consumption
Approx. 11 VA (supplied to the AC adapter)
Mass of product
Approx. 3.8 kg
*
1: With MID. , factory setting of the rate of the response adjustment, the stabilization time is approx. 3
seconds.
*
2: Factory setting
*
3: To use the counting mode or the percent mode, storing each mode as described in “9-9. Storing
Units” is required.