Specifications
65
Display Error code Description
EC, E00
Communications error
A protocol error occurred in communications. Confirm the format, baud
rate and parity.
EC, E01
Undefined command error
An undefined command was received. Confirm the command.
EC, E02
Not ready
A received command can not be processed.
e.g. The balance received a Q command, but not in the weighing
mode.
e.g. The balance received a Q command while processing a RE-ZERO
command. Adjust the delay time to transmit a command.
EC, E03
Timeout error
If the timeout parameter is set to "
t-Up1", the balance did not receive
the next character of a command within the time limit of one second.
Confirm the communication.
EC, E04
Excess characters error
The balance received excessive characters in a command.
Confirm the command.
EC, E06
Format error
A command includes incorrect data.
e.g. The data is numerically incorrect. Confirm the command.
EC, E07
Parameter setting error
The received data exceeds the range that the balance can accept.
Confirm the parameter range of the command.
Other errors
If the errors described above can not be released or other errors are
displayed, contact the local A&D dealer.
18-3. Other Display
The display ( ) of the password input is displayed when
using the password function and turning on the balance.
Enter a correct password. Refer to "
13. Password Function".
18-4. Asking For Repair
If the balance needs service or repair, contact the local A&D dealer.
The balance is a precision instrument. Use much care when handling the balance and observe the
following when transporting the balance.
Use the original packing material.
Remove the weighing pan from the main unit