Specifications
53
13. Password Function
The password function can be specified for an authorized user by using a password stored in the
balance by manager. While the password function is used, the balance can not be weighed without
inputting the correct password.
Password
: Four digits (4 x 4 x 4 x 4 = 256 outcomes) using MODE , SAMPLE , PRINT and
RE-ZERO keys.
At factory setting, the password function is not used.
The management of the password function is performed in "
Password function" of "9-3. Details Of
The Function Table".
13-1. Using Password Function
The password function can be switched between "Use" or "Not used" at "Password function (locfnc)"
of "
9-3. Details Of The Function Table".
1 Press and hold the
SAMPLE key in the weighing mode.
ba5fnc is displayed.
2 Press the
SAMPLE key several times until locfnc is
displayed.
3 Press the
PRINT key to display .
To cancel the operation, press the CAL key.
4 Press the
RE-ZERO key to display .
5 Press the
PRINT key to display where "No"
is selected.
6 "YES" and "No" can be switched with the
RE-ZERO key.
Press the RE-ZERO key to display where
"YES" is selected.
7 Press the
PRINT key to store the new setting. The password
function can now be used.
When turning on the balance, the password input is displayed.
Enter ke
y
Press and hold
Press several time
s
Exchan
g
e switch
"No" is selected
"Yes" is selected
Weighing mod
e
"OFF" is selected
"ON" is selected
Exchan
g
e switch
Enter ke
y
Enter ke
y