Specifications
52
12. Programmable-Unit
This is a programmable unit conversion function. It multiplies the weight data in grams by an arbitrary
coefficient set in the function table and displays the result.
The coefficient must be within the range between the minimum and maximum shown below. If the
coefficient set is beyond the range, an error is displayed and the balance returns to the coefficient
setting mode, prompting to enter an appropriate value. A coefficient of 1 was set at the factory.
Model Minimum coefficient Maximum coefficient
FZ-1200CT FZ-700CT FZ-500CT
FX-1200CT FX-700CT FX-500CT
0.000001 1000
Operation
1 Press and hold the SAMPLE key until ba5fnc of the function table is displayed.
2 Press the SAMPLE key several times to display j.
3 Press the PRINT key. The balance enters the mode to confirm or set the coefficient.
Confirming the coefficient
4 The current coefficient is displayed with the first digit blinking.
When it is not to be changed, press the CAL key and go to step 6.
When it is to be changed, press the RE-ZERO key and go
to step 5.
Setting the coefficient
5 Set the coefficient using the following keys.
SAMPLE key..........To select a digit to change the value.
The selected digit blinks.
RE-ZERO key.........To change the value.
MODE key .............To change the decimal point position.
Each time the switch is pressed, the decimal
point position changes as follows:
PRINT key..............To store the new setting, display end
and go to step 6.
CAL key.................To cancel the new setting and go to step 6.
Quitting the operation
6 The balance displays Unit . Press the CAL key to exit the
programmable-unit function and return to the weighing mode.
Using the function
Press the MODE key to select the programmable-unit (no display on the unit section). Perform
weighing as described in "5-1. Basic Operation (Gram Mode)". After weighing, the balance displays
the result (weight data in grams x coefficient).
Confir
m
0
.000001
00
.00001
000
.0001
0000
.001
00000
.01
000000
.1
0000001
Decimal point
position using
MODE key