Specifications

50
Weight data
Time
Remarks
Signature
End block
AD-8121 format
Manufacturer
Model
Serial number
ID number
Date
Time
Title block
General data format
A & D
MODEL FX-1200CT
S/N 012345678
ID ABCDEFG
DATE 2011/12/31
START
TIME 12:23:34
WT +102.456 g
WT +102.461 g
WT +102.462 g
WT +102.461 g
WT +102.461 g
WT +102.453 g
WT +102.471 g
WT +102.464 g
END
TIME 12:34:56
REMARKS
SIGNATURE
A & D<TERM>
MO DE L FX-1200CT<TERM>
S/ N 012345678<TERM>
ID ABCDEFG<TERM>
DA TE <TERM>
<TERM> *
ST AR T<TERM>
TI ME <TERM>
<TERM> *
WT +102.456 g<TERM>
WT +102.461 g<TERM>
WT +102.462 g<TERM>
WT +102.463 g<TERM>
WT +102.451 g<TERM>
WT +102.453 g<TERM>
WT +102.471 g<TERM>
WT +102.464 g<TERM>
<TERM>
EN D<TERM>
TI ME <TERM>
<TERM> *
REMARKS<TERM>
<TERM>
<TERM>
SI GN ATURE<TERM>
<TERM>
<TERM>
----------------<TERM>
<TERM>
<TERM>
<TERM>
C
R
L
F
C
R
C
R
L
F
Space, ASCII 20h
Terminator, , or
Carriage return, ASCII 0Dh
Line feed, ASCII 0Ah
Title Block And End Block
When a weight value is recorded as the GLP data, "Title block" and "End block" are inserted at the
beginning and at the end of a group of weight values, in the GLP report.
Note
To output the report to an AD-8121B , use MODE 3 of the AD-8121B.
Operation
1 With the weight data displayed, press and hold the PRINT key until S5tart is displayed.
The "Title block" is output.
2 The weight data is output according to the parameter setting of the data output mode.
3 Press and hold the
PRINT key until recend is displayed. The "End block" is output.
DP format (
type 1 of the function table) DP format (type 1 of the function table)
When the setting is "
info 1": When the setting is "info 2":
With FZ-CT series, the time and date are
added to output data.