Specifications
39
9-11. Percent Mode (%)
This is the mode to display the weight value in percentage compared to a 100% reference mass and
is used for target weighing or checking the sample variable.
Selecting The Percent Mode
1 Press the MODE key to select % (percent mode).
Storing The 100% Reference Mass
2 Press the SAMPLE key to enter the 100% reference mass
storing mode.
Even in the storing mode, pressing the
MODE key will switch to
the next mode.
3 Place a container on the weighing pan, if necessary. Press the
RE-ZERO
key to cancel the weight (tare). The balance displays
100 0 % .
4 Place the sample to be set as the 100% reference mass on the
pan or in the container.
5 Press the
PRINT key to store the reference mass. The balance
displays 100.00 % . (The decimal point position depends on the
reference value.)
Notes
If the balance judges that the mass of the sample is too light to be
used as a reference, it displays
lo . Do not use the sample.
The reference mass stored, even if the AC adapter is removed, is
maintained in non-volatile memory
6 Remove the sample.
Reading The Percentage
7 Place a sample to be compared to the reference mass on the pan.
The displayed percentage is based on the 100% reference mass.
While the stabilization indicator is on, pressing the
PRINT key will
output the weight value, using the RS-232C serial interface.
Note
Peripheral equipment, that is sold separately , such as a printer or
a personal computer is required.
9-12. Description Of The Item "Application Function"
Capacity Indicator ( apf 1)
In the weighing mode, the indicator displays the weight data relative to the weighing capacity in
percentage. (Zero =
0%, maximum capacity = 100%)
Statistical Calculation Mode ( apf 2)
The mode statistically calculates the weight data, and displays or outputs the results. For details, refer
to "6-5 Statistical Calculation Mode".
Weighing pa
n
C
ontaine
r
Sampl
e
100
%
Weighing result compared to
100% reference mass
Place sampl
e
to be weighed