Specifications

38
For a more accurate counting operation, use samples with a unit mass of at least ten or more times
the minimum weighing value of the balance. For example, when a balance with a mimimum weighing
value of 0.0001 g is used, the recommended sample unit mass is 0.001 g or more.
The sample unit mass stored, even if the AC adapter is removed, is maintained in non-volatile
memory.
Counting Operation
7 Place the samples to be counted on the pan.
While the stabilization indicator is on, pressing the
PRINT key will
output the weight value (number of objects), using the RS-232C
serial interface.
Note
Peripheral equipment, that is sold separately , such as a printer or
a personal computer is required.
Counting Mode Using The ACAI Function
The ACAI is a function that improves the accuracy of the unit mass
automatically by increasing the number of samples and averaging
the unit mass variable to minimize the weighing error, as the
counting process proceeds.
8 If a few more samples are added, the processing indicator turns on. To prevent an error, add three or
more. The processing indicator does not turn on if overloaded. Try to add the same number of
samples as displayed.
9 The balance re-calculates the unit mass while the processing indicator is blinking. Do not touch the
balance or samples on the pan until the processing indicator turns off.
10 Counting accuracy is improved when the processing indicator turns off.
Each time the above operation is performed, a more accurate unit mass will be obtained. There is no
definite upper limit of ACAI range for the number of samples exceeding 100. Try to add the same
number of samples as displayed.
11 Remove all the samples used in ACAI and proceed with the counting operation using the improved
unit mass.
Processing indicator