Specifications

30
DP (Dump Print) Format 5if type 1
This format is suitable for the peripheral equipment that prints the received data as is.
When an AD-8121B printer is used, set the printer to MODE 3.
This format consists of sixteen characters excluding the terminator.
A header of two characters indicates the balance condition. No overload header is used.
The polarity sign is placed before the data, with spaces in place of leading zeros, if the data is not zero
or overloaded.
The unit, consisting of three characters, follows the data.
KF Format 5if type 2
This is the Karl-Fischer moisture meter format and is used when the peripheral equipment can only
communicate using this format.
This format consists of fourteen characters excluding the terminator.
This format has no header characters.
The polarity sign is placed before the data, with spaces in place of leading zeros, if the data is not zero
or overloaded.
This format outputs the unit only for a stable value.
MT Format 5if type 3
This format is used when the peripheral equipment of other manufacturer is connected. Please note
that the connection is not guaranteed.
A header of two characters indicates the balance condition.
The polarity sign is used only for negative data.
The weight data uses spaces in place of the leading zeros.
The character length of this format changes dependent upon the unit
Terminator
WT +1.278 ct
C
R
L
F
Header Data
Unit
Stable header
Unstable header
Overload header
S
SD
SI
S 1.278ct
C
R
L
F
Header Unit Terminator
Data
+ 1.278 ct
C
R
L
F
Data Unit Terminator