Specifications
29
Key Mode C
When the PRINT key is pressed with the stabilization indictor turned on, the balance outputs the
weight data. When the stabilization indicator is not on, the balance waits for the indicator to turn on and
outputs the data. The balance blinks one time in this mode.
Required setting dout prt 5 Key mode C
Interval Output Mode
The weight data is periodically output. When the PRINT key is pressed, the balance starts to
output the weight data at a preset interval time. When the PRINT key is pressed again, the
balance stops outputting the weight data.
Example "For outputting the weight data periodically."
Required setting dout prt 6 Interval output mode
dout int Interval time
Caution
The balance may not transmit the data completely at the specified interval times and baud rate. Set
the baud rate higher.
9-6. Description Of The Item "Data Format"
A&D Standard Format 5if type 0
This format is used when the peripheral equipment can receive the A&D format. If an AD-8121B
printer is used, set the printer to MODE 1 or 2.
This format consists of fifteen characters excluding the terminator.
A header of two characters indicates the balance condition.
The polarity sign is placed before the data with the leading zeros. If the data is zero, the plus sign is
used.
The unit, consisting of three characters, follows the data.
Output example:
Print example:
Terminator
WT +1.278 ct
C
R
L
F
Header Data
Unit
Unit
DataHeader
ST,+0001.278 ct
C
R
L
F
Terminator