Specifications
26
Class Item and Parameter Description
0 Normal weighing mode
1 Capacity indicator
apf
Application function
2 Statistical calculation mode
0 Number of data, sum
1
Number of data, sum, maximum, minimum, average, range
(maximum-minimum)
2
Number of data, sum, maximum, minimum, average, range
(maximum-minimum), standard deviation, coefficient of variation
ap fnc
Application
5taf
Statistical function
mode output items
3
Number of data, sum, maximum, minimum, average, range
(maximum-minimum), standard deviation, coefficient of variation,
relative error
0 No used
locfnc
Password function
1 Used
When turning on the balance, the password input
( ) is displayed.
loc no.
Password
Refer to "13-2. Changing Password".
C5 in 1 *
Internal mass value correction: Method 1
C5 in 2 *
Internal mass value correction: Method 2
Refer to "7. Calibration".
Displayed only when the function
switch (internal mass value
correction switch) is set to 1.
: Factory setting Digit is a unit of minimum weighing value
*
: Only for the FZ-CT series
9-4. Description Of The Class "Environment, Display"
Condition ( Cond )
Cond 0 This parameter is for sensitive response to the fluctuation of a weight value. Used for powder
target weighing, weighing a very light sample or when quick response weighing is required.
After setting, the balance displays FAST .
Cond 2 This parameter is for stable weighing with slow response. Used to prevent a weight value
from drifting due to vibration or drafts. After setting, the balance displays SLOW .
Note
With "Hold function (Hold)" set to "ON (1)", this item is used to set the averaging time.
Stability Band Width ( 5t-b )
This item controls the width to regard a weight value as a stable value. When the fluctuation per
second is less than the parameter, the balance displays the stabilization indicator and outputs the data.
The parameter influences the "Auto print mode".
5t-b 0 This parameter is for sensitive response of the stabilization indicator. Used for exact weighing.
5t-b 2 This parameter ignores slight fluctuation of a weight value. Used to prevent a weight value
from drifting due to vibration or drafts.
Note
With "Hold function (Hold)" set to "ON (1)", this item is used to set the stabilization range.