Specifications
25
Class Item and Parameter Description
0 600 bps
1 1200 bps
2 2400 bps
3 4800 bps
4 9600 bps
bp5
Baud rate
5 19200 bps
0 7 bits, even
1 7 bits, odd
btpr
Data bit, parity bit
2 8 bits, none
0 CR LF
Crlf
Terminator
1 CR
CR: ASCII code 0Dh
LF: ASCII code 0Ah
0 A&D standard format
1 DP format
2 KF format
3 MT format
4 NU format
type
Data format
5 CSV format
Refer to "9-6. Description of the
Item "Data Format"".
0 No limit
t-Up
Timeout
1 1 second
Selects the wait time to receive a
command.
0 No output
5if
Serial
interface
erCd
AK, Error code
1 Output
AK:ASCII code 06h
mlt
Programmable-unit (Multi-unit)
Sets an arbitrary coefficient.
Available only when programmable-
unit mode is selected.
Unit
Unit
Refer to "9-8. Weighing Units".
id
ID number
Refer to "10. ID Number And GLP Report"
: Factory setting Digit is a unit of minimum weighing value
*
: Only for the FZ-CT series
Caution
The balance may not transmit the data completely at the specified refresh rate, depending on the
baud rate.
When a refresh rate of 20 times/second is used, set the baud rate to 4800 bps or higher.