Specifications

23
9-3. Details Of The Function Table
Class Item and Parameter Description
0 Fast response, sensitive value
FAST
1
MID.
Cond
Condition
2 Slow response, stable value
SLOW
With "Hold 1", sets the
averaging time.
0 Stable range is ±1 digit
1
5t-b
Stability band width
2 Stable range is ±3 digits
The stabilization indicator illuminates when
the display fluctuation is within the range.
With "
Hold
1
", sets the stabilization range.
0 OFF
Hold
Hold function
1 ON
Holds the display when stable in animal
mode. With "
Hold
1
",
HOLD
turns on.
0 OFF
1 Normal
2 Strong
trc
Zero tracking
3 Very strong
Keeps zero display by tracking zero drift.
0 Approx. 5 times/second
5pd
Display refresh rate
1 Approx. 10 times/second
Period to refresh the display
0 Point (.)
pnt
Decimal point
1 Comma (,)
Decimal point format
0 OFF
p-on
Auto display-ON
1 ON
Turns on the weighing mode display when
the AC adapter is connected.
0 OFF
poff
Auto display-OFF
1 ON (10 minutes)
Turns off the display after 10 minutes of
inactivity.
0 Displays
rng
Display at start
1 Does not display
Select whether or not to display the
minimum weighing value at weighing start.
0 Does not sound
ba5fnc
Environment
Display
beep
Beep
1 Sounds
Select whether or not to sound the beep
when operating on keys.
Cl adj *
Clock
Refer to "9-10. Clock and
calendar function".
Confirms and sets the time and date. The
time and date are added to output data.
0 No comparison
1
Comparison, excluding "near zero" when stable value or overloaded
2
Comparison, including "near zero" when stable value or overloaded
3 Continuous comparison, excluding "near zero"
Cp
Comp mode
4 Continuous comparison, including "near zero"
0 OFF
bep
-
LO buzzer
1 ON
0 OFF
bep-
OK buzzer
1 ON
0 OFF
Cp fnc
Comparator
bep
-
HI buzzer
1 ON
Cp Hi
Upper limit
Cp lo
Lower limit
Refer to "9-14. Comparator Function".
: Factory setting Digit is a unit of minimum weighing value
*
: Only for the FZ-CT series