Specifications

21
Operation
1 Press the ON:OFF key to turn off the display.
2 While pressing and holding the PRINT key and the SAMPLE key, press the ON:OFF key.
The balance displays p5 .
3 Press the PRINT key. Then the balance displays the function switches.
4 Set the switches using the following keys.
SAMPLE key
....
To select the switch to change the parameter. The selected switch blinks.
RE-ZERO key
...
To change the parameter of the switch selected.
0 : To inhibit changes. 1 : To permit changes
PRINT key
........
To store the new parameter and return to the weighing mode.
CAL key
...........
To cancel the operation ( Clr is displayed.)
To return to the weighing mode, press the CAL key once again.
8-2. Initializing The Balance
This function returns the following parameters to factory settings.
Calibration data
Function table
The sample unit mass value (counting mode), 100% reference mass value (percent mode)
External calibration weight
Function switch settings
Statistical data
Internal mass value (Only for the FZ-CT series)
Note
Be sure to calibrate the balance after initialization.
Operation
1 Press the ON:OFF key to turn off the display.
While pressing and holding the PRINT key and the SAMPLE key,
press the ON:OFF key. The balance displays p5 .
2 Press the
SAMPLE key to display Clr .
3 Press the PRINT key.
To cancel this operation, press the CAL key.
4 Press the RE-ZERO key.
5 Press the PRINT key to initialize the balance.
The balance will automatically return to the weighing mode.
Press and hold Pres