Specifications
19
7-7. Correcting The Internal Mass Value 2
(Only For The FZ-CT Series)
The FZ-CT series balance can correct the internal mass value using an external weight.
Method 2: Calibrate the balance using the external weight. The balance performs automatic
calibration using the internal mass, corrects the internal mass value and stores it in the
balance. The corrected mass value is maintained in non-volatile memory even if the AC
adapter is removed.
The usable calibration weights and the adjustable range are as follows:
Model Correction Reference Value Adjustable Range
FZ-1200CT 250 g, 200 g, 100 g, 50 g, 20 g
FZ-700CT 150 g, 100 g, 50 g, 20 g
FZ-500CT 100 g, 50 g, 20 g
-0.0150 g to +0.0150 g
The calibration weight in bold type: factory setting.
The calibration weight value can be adjusted within the range above.
Operation
1 With the factory setting, the balance can not correct the internal mass
value. Refer to "8. Function Switch And Initialization" and set the
internal mass value correction switch to 1.
2 In the weighing mode, press and hold the SAMPLE key to
display ba5fnc .
3 Press the SAMPLE key several times until C5 in 21 is
displayed. If C5 in 2 is not displayed, perform step 1.
4 Press the PRINT key to display Cal 0 .
Refer to "7-4. Calibration Using an External Weight", to
perform the calibration.
5 After calibration, remove the weight.
The balance displays . Press the PRINT key.
6 Cal. is displayed and the balance corrects the internal
mass value automatically
7 After correcting the internal mass value, the balance
displays Cal.
in
and performs the calibration
automatically using the corrected internal mass value.
8 The balance displays end and returns to the weighing
mode.
9 Confirm that the internal mass value has been
corrected properly by placing the external weight
used for the correction procedure on the
weighing pan.
If the internal mass value has not been corrected
properly, repeat the procedure to adjust the
correction value.
Calibration using an
external weight
Place the weigh
t
Press several time
s
Press and hold
Remove the weigh
t
A
djusting the internal mass valu
e
Calibrating with internal mas
s