Specifications

15
Press and hold
Release
GLP output
7-3. Calibration Test Using The Internal
Mass
(Only For The FZ-CT Series)
This function tests the balance accuracy using the internal mass.
Calibration test does not perform calibration.
This is available only when the "GLP output (info)" parameter is set to "1" or "2".
Operation
1 Connect the AC adapter and warm up the balance at least one hour.
2 Refer to "9. Function Tablel" to set "GLP output (info)"
to "1" or "2"
3 Press and hold the CAL key until CC in is
displayed, then release the key.
4 The balance measures the zero point.
Prevent vibration and drafts to affect the balance.
5 The measured zero point data is displayed.
6 The balance measures the internal mass.
Prevent vibration and drafts to affect the balance.
7 The value of the internal mass is displayed.
The normal range of the value is as follows:
Model The Internal Mass The Normal Range
FZ-1200CT 200.000 g
FZ-700CT
FZ-500CT
100.000 g
±0.002 g
8 If GLP output is set, glp is displayed, the
calibration test report is output to the RS-232C interfac.
Refer to "GLP output (info)" of the function table,
"10-2. GLP Report". end is displayed after the
calibration.
9 The balance automatically returns to the weighing mode.