Specifications
35
Operation
1 Press the ON:OFF key to turn off the display.
2 While pressing and holding the PRINT key and the SAMPLE key, press the
ON:OFF key. The balance displays p5 .
3 Press the PRINT key. Then the balance displays the function switches.
4 Set the switches using the following keys.
SAMPLE key To select the switch to change the parameter.
RE-ZERO key To change the parameter of the switch selected.
0: To inhibit changes. 1: To permit changes
PRINT key To store the new parameter and return to the weighing mode.
CAL key To cancel the operation ( Clr is displayed.) To return to the weighing
mode, press the CAL key once again.
8-2 Initializing the Balance
This function returns the following parameters to factory settings.
Calibration data
Function table
The sample unit mass value (counting mode), 100%
reference mass value (percent mode)
External calibration weight
Function switch settings
Statistical data
Note
Be sure to calibrate the balance after initialization.
Operation
1 Press the ON:OFF key to turn off the display.
2 While pressing and holding the
PRINT key and the
SAMPLE key, press the ON:OFF key.
The balance displays p5 .
3 Press the
SAMPLE key to display Clr .
4 Press the
PRINT key.
To cancel this operation, press the CAL key.
5 Press the
RE-ZERO key.
6 Press the
PRINT key to initialize the balance.
The balance will automatically return to the weighing mode.
With these keys
held down
Press the key