Specifications
25
5-5 Statistical Calculation Mode (Example of Use)
Here, as an example of use of the statistical calculation mode, mixing of
the multiple formulae such as medicine is described. The mixing
process is recorded using the balance and the printer.
In the example, the FX-3000i WP and the AD-8121B (dump print mode)
are connected using the RS-232C serial interface.
5-5-1 Getting Started
Changing the function table
Changes : To enable the statistical calculation mode
To enable “Zero after output”
Enabling the statistical calculation mode
1 Enter the function table menu.
Press and hold the
SAMPLE key until ba5fnc of the
function table is displayed, then release the key.
2 Select the application function.
Press the
SAMPLE key several times to display ap fnc .
Then, press the PRINT key to display °apf 0 .
3 Change the application function parameter to “2”.
Press the
RE-ZERO key to display apf 2 .
Press the PRINT key to confirm the change.
After end , ba5fnc is displayed.
Enabling “Zero after output”
4 Select “Zero after output”.
Press the
SAMPLE key several times to display dout .
Then, press the PRINT key to display °prt 0 , and press
the SAMPLE key several times to display °ar-d 0 .
5 Enable “Zero after output”.
Press the
RE-ZERO key to display ar-d 1 .
Then, press the PRINT key to confirm the change. After
end , 5if is displayed.
Returning to the weighing mode
6 Press the CAL key to return to the weighing mode.
Return to the weighing mode
Press
several times
Press and hold
Press
several times
Press
several times