Specifications
19
5-3 Percent Mode (%)
This is the mode to display the weight value in percentage compared to a 100% reference mass and is
used for target weighing or checking the sample variable.
Selecting the percent mode
1 Press the MODE key to select % (percent mode).
Storing the 100% reference mass
2 Press the SAMPLE key to enter the 100% reference mass
storing mode.
Even in the storing mode, pressing the
MODE key will switch
to the next mode.
3 Place a container on the weighing pan, if necessary.
Press the RE-ZERO key to cancel the weight (tare). The
balance displays 100 0 % .
4 Place the sample to be set as the 100% reference mass on the
pan or in the container.
5 Press the
PRINT key to store the reference mass. The
balance displays 100.00 % . (The decimal point position depends
on the reference value. The reference mass stored, even if the
AC adapter is removed, is maintained in non-volatile memory.)
Note
If the balance judges that the mass of the sample is too light to
be used as a reference, it displays lo . Do not use the sample.
6 Remove the sample.
Reading the percentage
7 Place a sample to be compared to the reference mass on the
pan. The displayed percentage is based on the 100% reference
mass.
While the stabilization indicator is on, pressing the
PRINT key
will output the weight value, using the RS-232C serial interface.
Note
Peripheral equipment, that is sold separately , such as a
printer or a personal computer is required.