Specifications

Weighing capacity
Minimum weighing value
Repeatability (Standard Deviation)
Linearity
Stabilization time (on FAST setting)
Sensitivity drift
(10°C to 30°C, when automatic self calibration is not used)
Operating temperature & humidity range
Calibration
Display refresh rate
Counting mode
Percent mode
Transmission
Water and dust proof specifications
Weighing pan dimensions
External dimensions
Power supply (AC adapter)
Net weight
Minimum unit mass
Number of samples
Minimum 100% reference mass
Minimum percentage display
GX-8K
8.1kg
0.01g
0.01g
±0.03g
±2ppm/°C
0.01g
1g
GX-8K2
8.1kg/2.1kg
0.1g/0.01g
0.05g/0.01g
±0.1g/±0.02g
±3ppm/°C
0.1g
10g
GX-12K
12kg
GX-10K
10.1kg
0.01g
0.01g
±0.03g
±2ppm/°C
0.01g
1g
GX-20K
21kg
0.1g
0.1g
±0.2g
±3ppm/°C
0.1g
10g
GX-30K
31kg
GX-32K
31kg/6.1kg
1g/0.1g
0.5g/0.1g
±1g/± 0.2g
±5ppm/°C
1g
100g
Approximately 1.5 seconds
5°C to 40°C, below 85%RH (no condensation)
Motor-driven Internal Calibration
5 times/second or 10 times a second
10, 25, 50 or 100pieces
0.01%, 0.1%, 1% (depends on reference mass stored)
RS-232C (provided as standard, with WinCT, our proprietary data transmission software)
Compliant with IP65
270(W) x 210(D) mm (10.62 x 8.26 inch)
300(W) x 355(D) x 111(H) mm (11.81 x 13.97 x 4.37 inch)
Confirm that the adapter type is correct for the local voltage and power receptacle
Approximately 9.3kg
1. Even if a heavy tare is used, the measurement can still be made within the fine range. (Smart range function) (GX-8K2/32K)
2. We recommend regular calibration and maintenance using an external calibration mass for more accurate linearity.
3. The internal mass is equal to (approximately) 860 grams.
4. The unit is equipped with an underhook for measuring density and weighing magnetic materials.
Options
GX-04K*
1
GX-06K*
1
GX-07K*
2
GXK-012
*
1
GX-04K, GX-06K, and the RS-232C (provided as standard)
cannot be used at the same time.
Water and dust proof specifications cannot be applied when
the GX-04K or GX-06K is installed.
*
2
Use the GX-07K optional water proof RS-232C cable in order to use the
standard RS-232C interface for water and dust proof (IP65) capabilities.
Comparator Output with Buzzer/
RS-232C/current loop output
Analog output/current loop output
Waterproof RS-232C cable (5m)
Animal weighing pan
Physical Dimensions (mm/inches)
Accessories
AD-8121B
AD-8118B
AD-8920
AD-8922
AD-1683EX
AD-1682
AX-KO1710-200
AX-SW-128
Dot Matrix Compact Printer
Universal Printer
Remote Display
Remote Controller
Static Eliminator
Rechargeable Battery Unit
(Maximum operating time: 8hours)
RS-232C Cable (25P-9P,2m)
Foot Switch
Specifications subject to change for improvement without notice.
...Clearly a Better Value
*
GX/GF-K-ADCC-03-PR3-05b05
GX-K
237 / 9.33
Weighing capacity
Minimum weighing value
Repeatability
(Standard Deviation)
Linearity
Stabilization time (on FAST setting)
Sensitivity drift (10°C to 30°C)
Operating temperature & humidity range
Display refresh rate
Counting mode
Percent mode
Transmission
Water and dust proof specifications
Weighing pan dimensions
External dimensions
Power supply (AC adapter)
Net weight
Minimum unit mass
Number of samples
Minimum 100% reference mass
Minimum percentage display
GF-8K
8.1kg
0.01g
0.01g
±0.03g
±2ppm/°C
0.01g
1g
GF-10K
10.1kg
0.01g
0.01g
±0.03g
±2ppm/°C
0.01g
1g
GF-8K2
8.1kg/2.1kg
0.1g/0.01g
0.05g/0.01g
±0.1g/±0.02g
±3ppm/°C
0.1g
10g
GF-12K
12kg
GF-20K
21kg
0.1g
0.1g
±0.2g
±3ppm/°C
0.1g
10g
GF-30K
31kg
GF-32K
31kg/6.1kg
1g/0.1g
0.5g/0.1g
±1g/± 0.2g
±5ppm/°C
1g
100g
Approximately 1.5 seconds
5°C to 40°C, below 85%RH (no condensation)
5 times/second or 10 times a second
10, 25, 50 or 100pieces
0.01%, 0.1%, 1% (depends on reference mass stored)
RS-232C (provided as standard, with WinCT, our proprietary data transmission software)
Compliant with IP65
270(W) x 210(D) mm (10.62 x 8.26 inch)
300(W) x 355(D) x 111(H) mm (11.81 x 13.97 x 4.37 inch)
Confirm that the adapter type is correct for the local voltage and power receptacle
Approximately 8.3kg
1. Even if a heavy tare is used, the measurement can still be made within the fine range. (Smart range function) (GF-8K2/32K)
2. The unit is equipped with an underhook for measuring density and weighing magnetic materials.
GF-K
AD-8121B
355 / 13.98
345 / 13.58
210 / 8.27
300 / 11.81
111 / 4.37
270/ 10.63
250 / 9.84
10 / 0.39
55/ 2.17
63 / 2.48