Instruction manual
C-12
CALF—Calibration Functions
ST: In case of the ST series, do not change the settings. If changed, weighing will not be performed properly.
Function #
Function name Range
Default
CALF-01 ST Weighing unit 2
0 None
1 g
2 kg
3 t
CALF-02 ST
Decimal point position 0
0 None
1 2 3 4 5
1
10
1
1 2 3 4.5
2
10
2
1 2 3.4 5
3
10
3
1 2.3 4 5
4
10
4
1.2 3 4 5
CALF-03 ST Minimum division 1, 2, 5, 10, 20, 50 (D) 1
Minimum division (D) for the weight. Input 1, 2, 5, 10, 20, or 50(decimal point ignored).
CALF-04 ST Capacity
0 – 800000
Decimal point ignored
10000
Capacity of the weighing. Weighing is allowed up to this setting + 9D(9 divisions). A weight beyond
this limit is an overflow and not displayed.
CALF-05 Zero range 0 - 30 2
A range of accepting "zero" from the [ZERO] key or Control I/O. Represented in terms of percent (%)
with respect to the weighing capacity, centering to the zero calibrated point in calibration. If this is set
to 2, "zero" can be accepted within a range of the zero calibrated point +/- 2 %.